TM 11-6625-2781-14&P-2
ERRATA (Cont'd)
Change " ... such as Anzac TDN-5350" to " ..... such as HP 11694A."
Changes supplement.
Change ANALYZER RANGE MHz to read: 0 - 110.
Replace entire Residual Responses test with paragraph 4-29 shown in this supplement.
Change entry for Residual Responses to the following:
Para.
Measurement
No.
Test Description
Units
Min
Actual
Max
Residual Responses
60-110 MHz; Residual Responses down
<-I10 dBm
dBm
-110
0-60 MHz; Residual Responses down
<-110 dBm
dBm
-110
1-10 MHz; Residual Responses down
<-110 dBm
dBm
-110
200 kHz-1 MHz; Residual Responses down
<-110 dBm
dBm
-110
20-200 kHz; Residual Responses down
<-95 dBm
dBm
- 95
Add the following:
WARNING
Adjustments described herein are performed with power supplied to the instrument while
protective covers are removed. Energy available at many points may, if contacted, result in
personal injury.
Add the following:
WARNING
Before proceeding to the next step, make sure the Display Section power is OFF.
B-6