TABLE 5-1
d. Adjust the GAIN control for a signal amplitude of
6 divisions.
e. Adjust the IF CENTER FREQ CAL for minimum signal
shift as the DISPERSION control is rotated between 10 MHz/
div and .2 MHz /div.
f. Position the IF feedthrough signal to the center of the
graticule with the Horizontal POSITION control.
g. Set the DISPERSION control to the .2 MHz/div posi-
tion.
h. Adjust the DISPERSION BAL for minimum signal shift
as the DISPERSION RANGE is switched between MHz and
kHz positions. Set the DISPERSION RANGE to kHz.
i. Adjust the IF CENTER FREQ CAL for minimum signal
shift as the DISPERSION control is switched between 100 kHz/
div and 1 kHz/div positions.
j. Check-There should be less than 2 major division
signal shift as the DISPERSION control is rotated down to
the 1 kHz/div position. The IF CENTER FREQ CAL adjust-
ment should not be against the stop.
k. Return the DISPERSION RANGE to MHz position and
10. Check Internal Triggering
the DISPERSION-COUPLED RESOLUTION control to 10 MHz/
div position.
a. Requirement-Sweep must trigger on 0.2 division signal
amplitude.
12. Check the Dispersion Accuracy of the MHz/
b. With 1 ms and 10 ns markers applied as in step 8,
DIV Ranges and the Range of the IF Center
adjust the Trigger LEVEL control for a triggered display on
Frequency Control
the INT position.
a. Requirement-Dispersion accuracy for the MHz/DIV
c. Decrease the amplitude of the displayed markers by
ranges is listed in Table 5-2. IF CENTER FREQ coarse con-
switching in 20 dB attenuation and adjusting the GAIN con-
trol range should equal or exceed + and - 25 MHz from
trol until sweep triggering can no longer be maintained
its centered (000) position. Dispersion accuracy and display
with optimum adjustment of the LEVEL control.
linearity must remain within the listed specifications of Table
5-2 to the + and 25 MHz positions of the control.
d. Check-Amplitude of the markers must be equal to
or less than 0.2 divisions (1 minor division).
b. Apply .1s and 10 ns time markers from the Time-Mark
Generator (Type 184) through a 20 dB attenuator to band
A RF INPUT connector.
c. Set the VERTICAL DISPLAY switch to LOG position.
a. Requirement-The center frequency of the IF band-
Adjust the GAIN control for a display amplitude of approxi-
pass with the IF CENTER FREQ controls centered must be
mately 6 divisions. Set the SOURCE switch to LINE and
adjustable to 200 MHz with the IF CENTER FREQ CAL
adjust the LEVEL control for a triggered display.
adjustment.
d. Center the IF CENTER FREQ controls.
b. Apply a calibrated 200 MHz signal from the Time-
e. Check the dispersion accuracy and linearity for each
Mark Generator (Type 184) to band B RF INPUT connector
MHz/DIV setting of the DISPERSION selector as listed in
through a 20 dB attenuator pad. (Signal input to the Type
491 should be less than -30 dBm to reduce the number
TION control or the IF CENTER FREQ control may be used
of spurious signals.)
to align the prime markers to the graticule divisions. As
the DISPERSION is decreased, the RESOLUTION control
c. Set the Type 491 front panel controls as follows:
should remain in the coupled position,
Position the trace to the
POSITION
f. Check the range, dispersion accuracy and linearity of
bottom graticule line
the IF CENTER FREQ control in the 2, 1, .5 and .2 MHz
I F CENTER FREQ
Centered (000)
positions of the DISPERSION selector.
Centered
FINE IF CENTER FREQ
Range of the coarse control should equal or exceed +
MHz/DIV
DISPERSION RANGE
and - 25 MHz from its centered position. It is checked
10 MHz/div
DISPERSION-COUPLED
by rotating the control to both extreme positions from center
RESOLUTION
and noting the frequency shift of the .1s or 10 MHz mark-
ers as the control is rotated. Dispersion accuracy and dis-
B
RF INPUT Selector
5-4