Fig. 6-43. Typical equipment setup to check
response flatness, sensitivity, frequency calibration and
phase lock operation for band C,
scales 7
through 8.
h. Center the IF CENTER FREQ controls ond the FINE RF
CENTER FREQ control then tune the signal to the center of
the screen with the RF CENTER FREQUENCY control (Hori-
zontal sweep must be centered.)
i. Check the dial accuracy as listed in Table 6-9. Must
equal or exceed (2 MHz + 1% of the dial reading).
j. As the dial accuracy is checked, depress the LOCK
CHECK button and check for phase lock beats. Check for a
phase lock display at the center and extreme frequency posi-
tions for each scale. Dial accuracy need only be checked for
scales 1, 2 and 4. The other scales are harmonic settings of
these fundamental ranges.
k. CHECK-Phase lock operation with an external refer-
ence frequency as follows:
1) Apply a 1 V peak to peak, 1 MHz signal, from the Con-
stant Amplitude Signal Generator (Type 191) to the REF
FREQ IN connector. Use a BNC T connector to apply the in-
put signal to the Type 491 to provide a convenient monitor-
must be measured at the REF FREQ IN connector. Turn the
INT REF FREQ control to the OFF or EXT REF FREQ IN posi-
tion.
Fig. 6-44. Signal to
noise ratio for measuring sensitivity.
6-42