33A1-13-170-1
Point of measurement
Normal indication
Isolating procedure
+430 volts dc.
If a close reading is not obtained, check V7C, C17 and C37, R84 and
R85, and L1.
Junction of C18D (C38C) and R68. (fig.
+230
volts dc.
If a close reading is not obtained, check C18D and R67,
+320 volts dc.
If a close reading is not obtained, check C17C, C17D, (C38C),
C18B, R49, R50, and R51.
+225 volts dc.
If the indication varies by more than a few volts, adjust R43. If
adjustment of R43 does not correct the reading, check V8, V9,
and V10 and associated circuit components.
7-3. Dc Resistance of Transformers and Coils
b. Do not use the resistance measurements as
The dc resistance data (table 73 below) is provided
the sole basis for discarding a transformer or coil
as an aid to troubleshooting (sec II). When using the
as defective. Because of rather broad winding
data, observe the following:
windings, check to see if abnormal operation is due to
values given in the table are average values. The
Figure 7-1. Spectrum Analyzer TS723A/U chassis,
top view, location of parts.